DICOM private fields not written correctly with custom dicomdict

조회 수: 9 (최근 30일)
Brian Keating
Brian Keating 2018년 8월 9일
편집: Jayanti 2025년 8월 21일
I've created a dicom dict text file that contains (among other things) a floating point private tag (9003,1038).
>> current_dicom_dict = dicomdict('get_current')
current_dicom_dict =
'/Users/brian/apps/copilot/share/dicom-dict-cp.txt'
>> !cat /Users/brian/apps/copilot/share/dicom-dict-cp.txt | tail -n 1
(9003,1038) FL DaysSincePriorStudy 1
The value of this tag is correctly set in the metadata struct:
>> metadata_out.Private_9003_1038
ans =
single
24.3231
I write the file with DICOMWRITE and read the file metadata back in with DICOMINFO:
dicomwrite(dat, 'test.dcm', metadata_out, 'CreateMode', 'copy', 'WritePrivate', 1);
metadata_in = dicominfo('test.dcm');
The metadata that I read back in failed to parse the private field correctly:
>> metadata_in.Private_9003_1038
ans =
4×1 uint8 column vector
176
149
194
65
Why is my private field not parsed correctly by DICOMINFO?

답변 (1개)

Jayanti
Jayanti 2025년 8월 21일
편집: Jayanti 2025년 8월 21일
Hi Brian,
It looks like the issue might be due to the fact that "dicominfo" is not using your custom DICOM dictionary when reading the file back.
So before calling "dicominfo", set your custom dictionary like below:
dicomdict('set', current_dicom_dict);
For more details please refer to the below offical documentation link on "dicomdict":

카테고리

Help CenterFile Exchange에서 Read and Write Image Data from Files에 대해 자세히 알아보기