Clarke Error Grid Analysis

버전 1.9.0.0 (27.1 KB) 작성자: Edgar Guevara
Clarke EGA quantifies the accuracy of glucose estimates generated by meters as compared to reference
다운로드 수: 4.4K
업데이트 날짜: 2013/4/1

라이선스 보기

The Clarke error grid approach is used to assess the clinical significance of differences between the glucose measurement technique under test and the venous blood glucose reference measurements. The method uses a Cartesian diagram, in which the values predicted by the technique under test are displayed on the y-axis, whereas the values received from the reference method are displayed on the x-axis. The diagonal represents the perfect agreement between the two, whereas the points below and above the line indicate, respectively, overestimation and underestimation of the actual values. Zone A (acceptable) represents the glucose values that deviate from the reference values by ±20% or are in the hypoglycemic range (<70 mg/dl), when the reference is also within the hypoglycemic range. The values within this range are clinically exact and are thus characterized by correct clinical treatment. Zone B (benign errors) is located above and below zone A; this zone represents those values that deviate from the reference values, which are incremented by 20. The values that fall within zones A and B are clinically acceptable, whereas the values included in areas C-E are potentially dangerous, and there is a possibility of making clinically significant mistakes. [1-4]

Syntax:

[total, percentage] = clarke(y,yp)

Inputs:
y = reference values (mg/dl)
yp = predicted/estimtated values (mg/dl)

Outputs:
total = total points per zone:
total(1) = zone A,
total(2) = zone B, and so on

percentage = percentage of data which fell in certain region:
percentage(1) = zone A,
percentage(2) = zone B, and so on.

Example:
load example_data.mat
[tot, per] = clarke(y,yp)

References:
[1] A. Maran et al., “Continuous subcutaneous glucose monitoring in diabetic patients: a multicenter analysis,” Diabetes Care, vol. 25, no. 2, pp. 347–352, Feb. 2002.
[2] B. P. Kovatchev et al. “Evaluating the accuracy of continuous glucose-monitoring sensors: continuous glucose-error grid analysis illustrated by TheraSense Freestyle Navigator data,” Diabetes Care, vol. 27, no. 8, pp. 1922–1928, Aug. 2004.
[3] E. Guevara and F. J. Gonzalez, “Prediction of Glucose Concentration by Impedance Phase Measurements,” in MEDICAL PHYSICS: Tenth Mexican Symposium on Medical Physics, Mexico City (Mexico), 2008, vol. 1032, pp. 259–261.
[4] E. Guevara and F. J. Gonzalez, “Joint optical-electrical technique for noninvasive glucose monitoring,” REVISTA MEXICANA DE FISICA, vol. 56, no. 5, pp. 430–434, Sep. 2010.

© Edgar Guevara Codina
codina@REMOVETHIScactus.iico.uaslp.mx
File Version 1.2
March 29 2013

Ver. 1.2 Statistics verified, fixed some errors in the display; thanks to Tim Ruchti from Hospira Inc. for the corrections
Ver. 1.1 corrected upper B-C boundary, lower B-C boundary slope ok; thanks to Steven Keith from BD Technologies for the corrections!
MATLAB ver. 7.10.0.499 (R2010a)

인용 양식

Edgar Guevara (2024). Clarke Error Grid Analysis (https://www.mathworks.com/matlabcentral/fileexchange/20545-clarke-error-grid-analysis), MATLAB Central File Exchange. 검색됨 .

MATLAB 릴리스 호환 정보
개발 환경: R2010a
모든 릴리스와 호환
플랫폼 호환성
Windows macOS Linux
카테고리
Help CenterMATLAB Answers에서 QSP, PKPD, and Systems Biology에 대해 자세히 알아보기

Community Treasure Hunt

Find the treasures in MATLAB Central and discover how the community can help you!

Start Hunting!
버전 게시됨 릴리스 정보
1.9.0.0

Ver. 1.2 Statistics verified, fixed some errors in the display; thanks to Tim Ruchti from Hospira Inc. for the corrections

1.8.0.0

Ver. 1.2 Statistics verified, fixed some errors in the display; thanks to Tim Ruchti from Hospira Inc. for the corrections.

1.2.0.0

Added some more references

1.1.0.0

Ver. 1.1 corrected upper B-C boundary, lower B-C boundary slope ok, thanks to Steven Keith from BD Technologies for the corrections!

1.0.0.0

Linefeeds corrected in description