Simulink Spectrum Analyzer channel power wrong (impedance?)

조회 수: 3 (최근 30일)
Daniel H
Daniel H 2016년 7월 25일
답변: Pertti Tolonen 2018년 12월 3일
Hi,
I have the following part if a SIMULINK model:
I assume the signals are voltages and I measure the signal power assuming 50-ohm matched system. Then the signal power is given by
P[dBm] = 10*log10(std(signal)^2 / 50 * 1000)
This is what I do in the small white blocks in front of the value displays. It can be seen that "RX in [dBm]" is -30dBm as I designed and expect.
Now I connect the SIMULINK Spectrum Analyzer in parallel and measure channel power. This is what I get:
-15.76dBm! This is far off. Now subtracting 10*log10(50) from this would give roughly -33dBm, close to -30 dBm.
Can it be that the Spectrum Analyzer block assumes 1 Ohm impedance? Can I set the input impedance to 50 Ohm?
Even then, the result would still be off by >3dB. Any idea for the discrepancies?

답변 (1개)

Pertti Tolonen
Pertti Tolonen 2018년 12월 3일
Yes: In order to get correct 50ohm dBm numbers, you must change the "Reference load" -setting in "Trace options". (n the screen grab above, the trace options menu is minimized.)
Struggled with the very same problem myself...

카테고리

Help CenterFile Exchange에서 RF Component Modeling에 대해 자세히 알아보기

Community Treasure Hunt

Find the treasures in MATLAB Central and discover how the community can help you!

Start Hunting!

Translated by